ksp-a100

Syringe pump KSP-A100

Syringe pump KSP-A100

Syringe pump KSP–A100 is a light weight, portable device designed for low-volume infusion. Equipped with LCD screen to display parameters like injection state, volume, mode, speed, syringe size, alarms, dosage, occlusion status, etc. for enhanced clinical efficiency. Inbuilt software based control system delivers higher levels of precision, safety and allows remote control mechanism to improve standardization, accuracy analysis of infusion data and reduce free–flow risks.

Download Catalog

Specifications

KVO Rate

0.1 – 5ml/h (steps : 0.1ml/h)

Injection Compliance

Adaptable for 5 syringe sizes

Preset Volume Range

0.1 ml – 1999.9 ml

Occlusion System

3-level alarming system

Mechanical Precision

≤ ±2 %

Injection Rate Range

50 ml Syringe 0.1 ml – 999.9 ml/h (step: 0.1 ml/h)
1000 ml – 1800 ml/h (step: 1 ml/h)
30 ml Syringe 0.1 ml – 900 ml/h (step: 0.1 ml/h)
20 ml Syringe 0.1 ml – 600 ml/h (step: 0.1 ml/h)
10 ml Syringe 0.1 ml – 300 ml/h (step: 0.1 ml/h)
5 ml Syringe 0.1 ml – 150 ml/h (step: 0.1 ml/h)

Purge Rate

50 ml Syringe 1800 ml/h (Bolus rate: 1200 ml/h)
30 ml Syringe 900 ml/h (Bolus rate: 600 ml/h)
20 ml Syringe 600 ml/h (Bolus rate: 400 ml/h)
10 ml Syringe 300 ml/h (Bolus rate: 200 ml/h)
5 ml Syringe 150 ml/h (Bolus rate: 100 ml/h)

Occlusion Alarm Threshold

High 800 mmHg ± 200 mmHg (106.7 kPa ± 26.7 kPa)
Medium 500 mmHg ± 100 mmHg (66.7 kPa ± 13.3 kPa)
Low 300 mmHg ± 100 mmHg (40.7 kPa ± 13.3 kPa)

Fuse

F1AL/250V (2 nos.)

Power Supply

100 V – 220 V, 50 Hz

Battery

Built–in Rechargeable Li–ion battery (11.1V)
Capacity ≥ 2000mAh ; 10 hours of charging
Works at 5ml/h after 4 hours of charging

Alarm

Near Injection finished, Preset dose injection finished, Occlusion, Syringe off, Syringe dislocated, Low Battery, On AC Supply.

Safety

Class I Type BF

Operation

Ambient Temperature 5°C to 40°C
Relative Humidity 20% – 90%
Atmospheric Pressure 86 kPa – 106 kPa

Storage

Ambient Temperature –30°C to 55°C
Relative Humidity ≤ 95%
Atmospheric Pressure 50 kPa – 106 kPa

Dimensions

282 x 210 x 148 mm

Weight

2 Kgs

Features

  • 4.3” color segment LCD
  • Multi injection modes – Volume/Time/Body
  • Adjustable speed, time, volume and dosage
  • Audio–visual alarms for critical events

Applications

    Used in general ward, ICU or post-operative care facilities in hospitals to gradually administer intravenous fluid drug dosage in controlled amounts at a predetermined rate and speed through syringe attached with a catheter for consistent and effective infusion therapy.