Intensive-Care Ventilator KIV-A100 is a high performance, closed-loop controlled, pneumatically-driven critical care ventilator with multiple ventilation modes. System features high resolution detachable touch screen with navigation knob to view monitored parameters, pulmonary mechanics waveform, lung volumes, flow-volume loops and three-level audio-visual alarm system with 360-degree panoramic alert light.
| Tidal Volume |
20 ml ─ 2500 ml |
| Configurations |
Trolley or Ceiling Pendant |
| Respiratory Rate |
1 bpm ─ 40 bpm (V-SIMV/P-SIMV/PRVC-SIMV) 4 bpm ─ 100 bpm ( VCV/PCV/PRVC) |
| Psupport |
0 kPa ─ 6 kPa |
| Pcontrol |
0.5 kPa ─ 6 kPa |
| Pulmonary Mechanics Waveform |
3 |
| Loops |
2 |
| Standard Ventilation Modes |
VCV (A/C) PCV (A/C) SIMV (VCV) SIMV(PCV) SIMV (PRVC) SPONT/CPAP NIV/CPAP NIV/PCV BIVENT |
| Combined Ventilation Modes |
SPONT+PS SIMV(VCV)+ PS SIMV(PCV)+PS SIMS(PRVC)+PS BIVENT+PS VCV+SIGH NIV/CPAP+PS |
| Inspiration Time |
0.1 s ─ 12 s |
| Inspiratory Pause Time |
0 s ─ 4 s |
| FiO2 Range |
21% ─ 100% |
| Trigger Sensitivity |
Pressure : - 2 kPa to 0 kPa (above PEEP) Flow : 0.5 LPM ─ 20 LPM (base flow) |
| PEEP |
0 kPa ─ 5 kPa |
| Suction |
< 2 min |
| Nebulizer |
30 mins |
| Pressure Values |
Pplat / Ppeak / Pmean / PEEP |
| Volume / Flow values |
VTI / VTE / MV / MVspont |
| Time Values |
Ftotal / Fspont / I : E |
| Oxygen Monitoring |
Without O2 sensor ( O2 sensor optional ) |
| Gas Supply |
O2, Air (0.28 MPa – 6 MPa) |
| Real-time Curves |
Pressure-Time Flow-Time Volume-Time Pressure-Volume loop Flow-Volume loop |
| Alarm |
Upper and Lower limits : MV, Paw, PEEP, Airway Pressure continuous high, VTE, Fspont high Other Conditions : Tapnea, ,Power Fail, Nebulizer, Battery low, Battery exhausted, Air supply failure, O2 supply failure, Air and O2 supply failure, Fan block, Pipe block, Circuit disconnected |
| Display |
12.1” TFT LCD touch screen |
| Internal Battery |
> 1 hour |
| Maximum Security Pressure |
≤ 8 kPa |
| Compliance |
≤ 4 ml / 100 Pa |
| Noise |
≤ 65 dB (A) |
| Communication |
RS232, VGA, Nurse Call |
| Operational Temperature |
5 °C to 40 °C |
| Relative Humidity |
≤ 90 % (non-condensing) |
| Atmospheric Pressure |
50 kPa – 106 kPa |
| Power Supply |
110 V – 240 V (50 Hz / 60 Hz), 65 VA |
| Dimensions |
328 x 410 x 410 mm |
| Weight |
20 Kgs |
Features
Applications
Used in operating rooms, surgical care wards, intensive care unit, and other departments for pulmonary life-support during acute lung injury or respiratory distress which result in inadequate oxygenation or ventilation inability to protect the airway fluids.